Product Description

3 main tests can be performed using this set

  • Photometric accuracy: Users can use the 3 density filters and plot a standard curve, then check the R value. The closer the R value is to 1, the more accurate the photometric accuracy of the spectrophotometer.

  • Wavelength Accuracy: Users can scan the Holmium and Didymium filter and match all peaks as shown in the test report provided along.

  • Stray Light: 220 nm filter is to test stray light in the lower UV range. 340 nm filter is to test stray light in higher UV range. The absorption in both cases should be equal to or greater than 3.

#1 - 10%T filter
#2 - 20%T filter
#3 - 30%T filter
#4 - 220 nm stray light filter
#5 - 360 nm stray light filter
#6 - Holmium filter (wavelength accuracy test) #7 - Didymium filter (wavelength accuracy test) #8 - BLANK filter

Accuracy: +/-0.2 nm or +/-1%

Each part is mounted in a standard 10 mm cuvette metal holder
Operating temperature: 21°C to 25°C