Product Description

The Labmart LMCM20H Conductivity / TDS Meter is an auto-ranging microprocessor-based laboratory meter designed for measurement of conductivity, Total Dissolved Solids (TDS), and temperature in liquids. It features automatic conductivity/TDS ranging, up to 3-point calibration, automatic temperature compensation, selectable cell constants, automatic endpoint detection, USB connectivity and memory for up to 100 readings. 

Product Overview

The Labmart LMCM20H Benchtop Conductivity / TDS Meter is designed for laboratory measurement of the electrical conductivity and Total Dissolved Solids of liquid samples, with an additional temperature measurement mode.

The instrument uses automatic ranging for conductivity and TDS measurements and supports conductivity calibration at up to three points. Calibration standards can be selected from 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm, with any three available for calibration.

Automatic or manual temperature compensation can be selected, while the cell constant can be configured to K = 0.1, 1 or 10. A large white-backlit LCD, automatic read function, 100-reading memory and USB interface support convenient laboratory operation and data handling.

Key Features and Benefits

  • Conductivity, TDS & Temperature Modes – Provides three measurement functions in one benchtop instrument.
  • Wide Auto-Ranging Conductivity Measurement – Automatically selects the appropriate measurement range across the specified conductivity ranges.
  • Up to 3-Point Calibration – Supports multi-point calibration using selectable conductivity standards.
  • Automatic Conductivity Standard Recognition – Helps simplify calibration by automatically recognizing supported conductivity standards.
  • Selectable Cell Constant – K values of 0.1, 1 and 10 can be selected according to measurement requirements.
  • Automatic Temperature Compensation – Supports manual or automatic compensation from 0 to 100°C.
  • Automatic Endpoint Detection – The auto-read function senses and locks the measurement endpoint.
  • TDS Factor Adjustment – TDS factor can be configured from 0.1 to 1.0, with 0.5 as the default.
  • 100-Reading Memory – Stores and recalls up to 100 readings.
  • USB Interface – Provides USB connectivity.
  • Large Backlit LCD – Provides clear viewing of measurement information in laboratory environments.

4. Technical Specifications

Specification Details
Product Type Conductivity / TDS Meter
Model LMCM20H
Brand Labmart
Source / Reference Manufacturer LABMAN
HSN Code 90275090
Measurement Modes Conductivity, TDS, Temperature
Conductivity Range 0–20.0, 200.0, 2000 µS/cm; 20.00, 200.0 mS/cm
Conductivity Ranging 5 auto ranges
Conductivity Resolution 0.01, 0.1, 1
Conductivity Accuracy ±0.5% F.S.
Calibration Points Up to 3 points
Calibration Solutions 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm; any three
TDS Range 0–10 ppt; maximum 200 ppt depending on factor setting and conductivity range
TDS Accuracy +1% F.S.
TDS Factor 0.1 to 1.0; default 0.5
Temperature Range 0 to 105°C / 32 to 221°F
Temperature Resolution 0.1°C / 0.1°F
Temperature Accuracy ±0.5°C / ±0.9°F
Temperature Compensation 0 to 100°C, manual or automatic
Cell Constant Options K = 0.1, 1, 10
Normalization Temperature 25°C
Memory Up to 100 readings
Reset Function Yes
Interface USB
Connector 6-pin Mini Plug
Operating Temperature 10 to 40°C
Relative Humidity <80%
Display LCD with white backlight
Display Size 25 × 100 mm
Power Supply DC adaptor, AC 220 V, 50/60 Hz
Packing Dimensions (L × W × H) 310 × 255 × 200 mm
Gross Weight 2 kg
Standard Accessories Main instrument, stand, adaptor, ATC probe, conductivity cell (1K), conductivity solution, instruction manual, test certificate

The specifications above are taken from the product specification table on page 3 of the supplied catalogue.

Note: The supplied source specifies TDS accuracy as +1% F.S.; this value has been preserved rather than interpreted or altered.


5. Applications

The LMCM20H is suitable for conductivity and TDS measurement applications such as:

  • Water quality testing
  • Laboratory water analysis
  • Conductivity testing of liquid samples
  • TDS measurement
  • Chemical solution testing
  • Environmental laboratory testing
  • Quality control testing
  • Research and development
  • Educational laboratory experiments
  • Process and formulation checks
  • General laboratory liquid analysis

6. Suitable For

  • Analytical laboratories
  • Research laboratories
  • Quality control laboratories
  • Water testing laboratories
  • Environmental laboratories
  • Chemical laboratories
  • Pharmaceutical laboratories
  • Food and beverage laboratories
  • Educational institutions
  • Industrial laboratories
  • R&D departments

7. FAQs

1. What does the Labmart LMCM20H measure?

The LMCM20H measures conductivity, Total Dissolved Solids (TDS), and temperature.

2. What is the conductivity range of the LMCM20H?

The specified conductivity ranges are 0–20.0, 200.0 and 2000 µS/cm, and 20.00 and 200.0 mS/cm, with five automatic ranges.

3. Does the LMCM20H have automatic ranging?

Yes. The instrument provides automatic ranging for conductivity and TDS measurements.

4. How many calibration points are available?

The LMCM20H supports up to 3-point calibration.

5. Which conductivity calibration standards are supported?

The specified calibration solutions are 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm, with any three usable for calibration.

6. What is the TDS measurement range?

The specified TDS range is 0–10 ppt, with a maximum of 200 ppt depending on the factor setting and conductivity range.

7. Can the TDS factor be adjusted?

Yes. The TDS factor can be set from 0.1 to 1.0, with 0.5 as the default setting.

8. Does the meter provide automatic temperature compensation?

Yes. Temperature compensation is available from 0 to 100°C, with manual or automatic operation.

9. What cell constants are supported?

The instrument provides selectable cell constants of K = 0.1, 1 and 10.

10. How many readings can be stored?

The internal memory can store and recall up to 100 readings.

11. What accessories are included?

The standard accessories include the main instrument, stand, adaptor, ATC probe, 1K conductivity cell, conductivity solution, instruction manual and test certificate.

12. Does the LMCM20H have USB connectivity?

Yes. The specified interface is USB.