MULTI-POINT PUSH-BUTTON CALIBRATION
The meter has multi-point calibration mode. From setup menu, choose calibration and immerse electrode into indicated calibration solution, press ENTER key and meter is automatically calibrated.
AUTO-BUFFER RECOGNITION
During calibration, if the calibration solution or selected value does not match the meter's indicated value, an ERROR message is displayed.
VIEWING OF ELECTRODE SLOPE
After calibration, the percentage slope of the electrode is displayed before the meter returns to measurement mode. Per the slope value, user decides whether to change the electrode
AUTOMATIC TEMPERATURE COMPENSATION
By using temperature probe
AUTO-RANGING FOR CONDUCTIVITY
The meter has auto-range functions for measuring conductivity and total dissolved solids (TDS). Upon immersing conductivity sensor into sample solution, the screen immediately displays the value in appropriate measurement range for accuracy,
ADJUSTABLE TEMPERATURE COEFFICIENT
Various ions in solution have different temperature coefficients. LABMAN-LMMP30. can set the temperature coefficient between 0 and 3.9% in degrees centigrade for optimum accuracy
ADJUSTABLE CELL CONSTANT
The meter has three cell constants for measuring requirements. Once a cell constant has its calibration set, it is saved and does not require re-calibration when switching meter
MESSAGE INDICATOR
LABMAN-LMMP30 displays helpful messages like an on-screen manual for easy operation of this meter and selection of the various options
SAVE AND COMMUNICATION INTERFACE
The meter all has real-time date/clock stamps. The memory stores up to 99 discrete data sets. Press the SAVE key to store the current measured values and it automatically generates a serial number with the date and time. This data can then be sent to a computer via the USB port
METER INCLUDES : pH Electrode, Conductivity cell (1.0), pH buffers, temperature probe, Electrode Stand, Power Adaptor & Instruction Manual.
Model No. | LMMP30 |
pH Range | -2.000 to I6.000pH |
pH Resolution | ±0.001pH |
pH Accuracy | ±0.002pH |
Calibration Points | 2 points, NIST(pH4.00/6.86/9.18) or USA(pH4.01/7.00/10.01] |
mV/ORP Range | 0 to ±1999.9mV |
mV/ORP Accuracy | ±0.2mV |
Temperature Compensation | 0.0 to 100.0°C, Manual or Automatic |
Conductivity Range | 0 to 1.999,19.99,199.9,1999 µ S, 19.99, 200.0mS |
Conductivity Accuracy | ± 0.5% F.S |
Calibration Points | 1 point per range |
Temperature Compensation | 0 to 50°C, Manual or Automatic |
Temperature Coefficient | 0 to 3.9% per°C |
Cell Constant | 0.1,1.0,10 adjustable |
TDS Range | 0 to 100 ppt, Max.200 ppt |
TDS Accuracy | +1% F.S |
Salinity Range | 0 to 50 ppt, 0.00 to 5.00% |
Accuracy | ±0.5ppt, ±0.05% |
Resistivity Range and Accuracy | 0 to 100M, W± 1%F.S |
Temperature Range | 0.0 to 100.0°C |
Memory & Output | 99 data sets with date/ time stamping, USB |
Power supply | DC 9 V, using AC Adapter 220 VAC, 50 Hz |
Brand | Labman |
---|